Skip to main content

Table 1 CTF Sensitivity

From: Error analysis in the determination of the electron microscopical contrast transfer function parameters from experimental power Spectra

V

61.23 (100.00)

s m

0.4890 (100.00)

V

0.2901 (100.00)

Δf M

37.93 (61.95)

s M

0.4613 (94.34)

Δf M

0.2507 (86.42)

Δf m

37.45 (61.16)

b

0.2447 (50.04)

Δf m

0.2497 (86.07)

C a

6.132 (10.01)

K G

0.2441 (49.92)

Q 0

0.0331 (11.41)

ΔV/V

5.32 (8.69)

V

0.0371 (7.59)

θ

0.0053 (1.83)

θ

1.51 (2.47)

Δf m

0.0288 (5.89)

T m

0.0010 (0.34)

Q 0

0.96 (1.57)

Δf M

0.0285 (5.83)

C s

0.0001 (0.03)

α

0.80 (1.31)

K

0.0096 (1.96)

ΔV/V

NA

C s

0.78 (1.27)

C a

0.0061 (1.25)

C a

NA

K

0.72 (1.18)

ΔV/V

0.0058 (1.19)

K

NA

ΔR

0.24 (0.39)

K s

0.0037 (0.76)

α

NA

ΔF

0.03 (0.05)

C m

0.0025 (0.51)

ΔR

NA

T m

NA

K g

0.0022 (0.45)

ΔF

NA

  

G M

0.0021 (0.43)

  
  

C M

0.0019 (0.39)

  
  

G m

0.0018 (0.37)

  
  

Q 0

0.0015 (0.31)

  
  

θ G

0.0013 (0.27)

  
  

c m

0.0013 (0.27)

  
  

α

0.0011 (0.22)

  
  

g M

0.0010 (0.20)

  
  

g m

0.0008 (0.16)

  
  

θ

0.0007 (0.14)

  
  

c M

0.0004 (0.08)

  
  

ΔR

0.0002 (0.04)

  
  

θ s

0.0002 (0.04)

  
  

C s

0.0001 (0.02)

  
  

θ g

0.0001 (0.02)

  
  

ΔF

NA

  
  

T m

NA

  
  1. Each column represents the average error sensitivity of the CTF (S H ), the theoretical PSD (S PSD ), and the first zero of the CTF (S zero ). Columns are sorted by descending sensitivity. In parenthesis is shown the percentage of the corresponding sensitivity with respect to the maximum sensitivity in that column. Those entries marked as NA mean that the parameter is not affecting the measured quantity.